Model | LIS-AM | LIS-ACM | LIS-A5CM |
Feature | 190L, Touch screen | 190L, Touch screen, with CO2 function | 190L, Touch Screen, with co2 function, Amplitude 50 |
Shaker Size | 550*470mm |
Number of flask clamps / stickers | Flask clamp: 50ml*72/100ml*30/150ml*30/200ml*30/250ml*30/500ml*17/1000ml*12/2000ml*8/5000ml (Thomson)*4 Adhesive pad: 50ml*72/100ml*53/150ml*41/200ml*32/250ml*30/500ml*20/1000ml*12/2000ml*8/5000ml (Thomson)*4 |
Capacity | 190L |
Maximum load | 10kg |
Temperature control range | Below room temperature 18 ℃ -60 ℃ (max. 4 ℃) |
Temperature control accuracy | +/-0.1℃ |
Temperature uniformity | +/-0.5 ℃(@37 ℃) |
Rotation speed | 0(stop),30-300rpm |
Rotation accuracy | +/-1 rpm |
Magnetic drive | YES |
Amplitude | 26mm | 50mm |
CO2 sensor | - | IR (infrared) |
CO2 concentration control range | - | 0-20% |
CO2 concentration control accuracy | - | +/-0.1% |
Temperature recovery time after opening door after 30S restart | - | ≤6 min (@37 ℃) |
CO2 concentration recovery time after 30S restart | - | ≤6 min (@5%) |
Humidity type | - | Natural humidification |
Humidity | - | 95% |
Programmable number of segments | 12 segments |
Illumination | Two-way lighting |
Sterilize | UV sterilization |
Door opening method | Left open |
Set parameter protection | password protection |
Curve display type | Temperature curve, speed curve, | Temperature curve, speed curve, CO2 concentration curve, humidity curve |
Display method | Capacitive, 7-inch color touch screen |
Timing range | 0-999 hours and 59 minutes |
USB flash drive export data | YES |
Days of operation parameters can be checked on the screen | Within 7 days |
Run data storage cycle | More than 912 days |
Operation data storage interval | 1 minute |
Data storage content | Running time, temperature parameters, speed parameters, running status | Running time, temperature parameters, speed parameters, humidity parameters, running status |
Automatic defrost | YES |
Turn off screen display automatically | YES |
Auto lock screen | YES |
Third shake plate height | 1315mm |
Rocker with track | YES |
Noise level | ≤55Db(When cooling) |
Washable | YES |
Drive | Magnetic drive |
Interface language | English |
Automatic call reply function | YES |
Heating power | ≤600W |
Cooling power | 70-180W |
Power | ≤730W | ≤780W |
Power supply | AC 220V+/-22V,50HZ/60HZ |
Active humidification (optional) | - | 10% -95% above ambient humidity(optional ) |
WIFI mobile monitoring (optional) | Optional |
Computer monitoring (optional) | Optional |
Net weight | 170kg |
Inner cavity working height | 303mm |
Dimensions | 970*780*(530+80 High feet) mm |